Performance

FLOW RATE RANGE From 1 to 1000 ml/h in normal mode.
From 0,1 to 100 ml/h in micro-infusion mode.
FLOW RATE ACCURACY ± 5%  with recommended sets.
PROGRAMMING MODES Rate only, rate volume, volume/ time, ramp up/  ramp down, primary/ secondary, sequential, bolus, micro infusion.
VOLUME LIMIT From 1 to 9999 ml in normal mode, from 0,1to 999,9 in micro infusion mode.
INFUSION TIME Settable from 1minute to 96 hours.
KVO RATE 3 ml/h, programmable.
AIR DETECTION Set  at  250  µI over a period of 15 minutes.
OCS: OCCLUSIVITY CHECK SYSTEM OCS patented system is the only auto-test which checks the good operating of the  pump in association with its set and thus prevents all risk of free-flow.
FREE-FLOW PROTECTION Control of occlusivity of pumping system toward set, automatic safety clamp integrated into set.
DPS: DYNAMIC PRESSURE SYSTEM Line disconnection alarm, indication of pressure increase and decrease in the  infusion line.
OCCLUSION DETECTIONS Upstream & downstream.
OCCLUSION THRESHOLD Set at 750 mmHg, settable from 100 to 900  mmHg.
AUTOMATIC REDUCTION OF BOLUS
AFTER RELEASE OF AN OCCLUSION
Bolus < 0,2 ml, whatever the rate.
KEYBOARD LOCK Programmable in configuration.
DROP SENSOR Use in option for an empty bottle detection.
HISTORY MODULE Up to 768 dated events.
ALARMS AND PREALARMS Infusion control: set in place, volume limit reached, empty bottle, air in line,
upstream occlusion, downstream occlusion, set disconnection, end of pause, door open. Device control: discharged battery, low battery, interface failure, technical malfunction. Maintenance: periodic servicing.

Technical Specifications

PUMPING SYSTEM Second generation linear wave peristaltic pumping system.
DIMENSIONS/ WEIGHT H 10,5 x W 31,5 x D 13 cm/~ 2,4kg.
EXTERNAL SUPPLY 7,15 Vdc (with the Base or the Mainy Mod).
BATTERY/ BATTERY LIFE Lead acid battery  – 6 V 1.1to 1.3 Ah/   2 hours 125 ml/h. Display of real battery life in hours and minutes.
COMPLIANCE, STANDARDS Complies with EN 60601·1, EN 60601·1·2 and EN 60601·2·24.
CE 0459 Markino in compliance with the EEC 93/42 European Medical Device Directive.
WATERPROOFNESS Protection against splashing liquid: IP 34.
ELECTRICAL COMPLIANCE Protection against leakage current: CF type. Protection against electric shocks: class II.