Search

In Stock

CASMED 740 MAX NIBP Patient Vital Signs Monitor

NIBP

Technique: Oscillometric (MAXNIBP® Technology)
Patient Range: Adult – Neonate

Measurement Range

Systolic: 30-255 mmHg
Diastolic: 15-220 mmHg
MAP: 20-235 mmHg
Pulse Rate Range: 30-240 BPM

Accuracy

Blood Pressure: ± 5 mmHg standard deviation no greater than 8 mmHg
Pulse Rate: ± 2% or ±2 BPM, whichever is greater

Oximetry (Options) Masimo SET®

SpO2 Range: 0-100%

Nonin®

SpO2 Range: 0-100%

SpO2 Accuracy:

70-100%, ± 2 digits (1 S.D.) Adult
70-100%, ± 3 digits (1 S.D.) Neonate
Pulse Rate Range: 18-240 BPM
Pulse Rate Accuracy: ± 3% or ±1 digit, whichever is greater

Nellcor® OxiMax®

Temperature (Optional) Welch Allyn SureTemp® Plus

Temperature Range: 29 °C to 42 °C (84.2 °F to 107.6 °F)
Accuracy: ± 0.1 °C (± 0.2 °F) Meets or exceeds ASTM Standards

Control Panel

Display: LED display of measurement results, instructions, troubleshooting messages, and signal strength bar
Parameters Displayed: Systolic Pressure, Diastolic Pressure, Mean Arterial Pressure (MAP), Pulse Rate,
%SpO2, Temperature (°C or °F)

Patient Alarms

Adjustable high and low alarms for Pulse Rate, SpO2, Systolic Pressure, Diastolic Pressure and Mean Arterial Pressure

Operating Modes

History: Review of previous measurements
NIBP: Manual, STAT or Automatic (at preset intervals)
%SpO2: Continuous Monitoring
Temperature: Predictive or Continuous

Power

Source: External line or internal battery
AC Power: 100-240 VAC, 50-60 Hz, 0.5A;
T1.25A L250V fuse rating
DC Power (EMS Option): 12-30 VDC Battery: NiMH battery pack (removable) Charge Time: 4 hours
Operating Time: Minimum 4 hours of continuous operation
Leakage Current: 100 µA (maximum)

Safety Limits

145 mmHg (Neonate)
If measurement time exceeds 120 seconds (Adults), 90 seconds (Neonate)
If safety timer detects microprocessor failure

Operating Environment

Operating Temperature: 0° C to 50° C (32° F to 122° F)
Storage Temperature: -20° C to 65° C (-4° F to 149° F)
Humidity: 15 – 95%, non-condensing
Altitude: Up to 3048 meters (10,000 feet)
if stored or used outside temperature and humidity ranges.

Physical Dimensions & Weight (Base Unit)

H x W x D: 17.0 cm x 21.5 cm x 7.5 cm (6.75 in x 8.5 in x 3.0 in)
Weight: 1.4 kgs (3 lbs)

Standards

ANSI/AAMI SP-10
EN 60601-1, EN 60601-1-2, EN 865, EN 60601-2-30,
EN 60601-2-49
ETL Listed – UL 60601, CAN/CSA C22.2 No. 601.1
CE Marking according to Directive 93/42/EEC

Optional Accessories

Infrared printer RS-232 interface
Swiveled hard mount (for ambulance applications) Roll stand and basket
Protective carrying case

Condition

Service Manual

Download

Back to Top
Product has been added to your cart